Document EDFA max-power characterisation; restore safe pump clamp
Swept pump-1 to find real output: 1 W (30.1 dBm) at 4000 mA, 1.22 W at 4500 mA still unsaturated. External Thorlabs meter read 1.05 W vs 1.22 W internal PD -> telemetry is ~0.5-0.6 dB optimistic, path-dependent; treat amp/output*_power_dbm as indicative, not calibrated. - README: full pump-vs-output table, PD-vs-meter calibration note, >1 W output-connector warning, ramp/overshoot guidance - config.h: PUMP_CURRENT_MAX_MA back to 4000 mA (1 W rating) after the supervised over-rating sweep; added get_thresholds probe to agiltron driver
This commit is contained in:
+4
-1
@@ -26,7 +26,10 @@
|
||||
// HA number sliders are capped here. The hardware allows 8000 mA pump, but
|
||||
// the lab notes report ~2500 mA pump-1 -> ~27 dBm (~515 mW). Keep a margin.
|
||||
#define SEED_CURRENT_MAX_MA 300 // CTL200 ilmax default is 1000; seed DFB is low-power
|
||||
#define PUMP_CURRENT_MAX_MA 3000 // firmware clamps amp pump-1 to this
|
||||
// Pump-1 clamp = the 1 W rated operating point (~4000 mA, measured 2026-06-17).
|
||||
// Raise deliberately only for supervised over-rating characterisation; it is a
|
||||
// guardrail against accidental over-drive, not a safety interlock.
|
||||
#define PUMP_CURRENT_MAX_MA 4000
|
||||
|
||||
// ── Serial links to the instruments (wired in later increments) ─────
|
||||
// UART0 stays the USB console. The classic ESP32 has 3 hardware UARTs.
|
||||
|
||||