Swept pump-1 to find real output: 1 W (30.1 dBm) at 4000 mA, 1.22 W at
4500 mA still unsaturated. External Thorlabs meter read 1.05 W vs 1.22 W
internal PD -> telemetry is ~0.5-0.6 dB optimistic, path-dependent; treat
amp/output*_power_dbm as indicative, not calibrated.
- README: full pump-vs-output table, PD-vs-meter calibration note, >1 W
output-connector warning, ramp/overshoot guidance
- config.h: PUMP_CURRENT_MAX_MA back to 4000 mA (1 W rating) after the
supervised over-rating sweep; added get_thresholds probe to agiltron driver